It is a culture device integrating a multi-gas incubator that cultures fertilized eggs and a microscopic system
There is no stress on fertilized eggs because it is unnecessary to export fertilized eggs from the incubator at the time of observation
Features
① Stable culture environment (temperature).It is specially designed for 9 chamber dryers. Each Independent temperature sensor in the chamber Control the special heater with precision Sample temperature control, stable cultureWe provide the environment.
② Stable culture environment (gas). Adopted AGM (ASTEC Gas Mixing) system with proven track record in ASTEC dry incubator. From the built-in mix gas mixing system, supply to each chamber in a circulating manner.
③ You can freely shoot and observe specimens of the best class level 108 position (9 dishes × 12 well).
④ Built-in dedicated PC installed. We installed a built-in PC dedicated to the main body. The 15.6-inch touch screen further enhances operability and visibility, ensuring reliable communication with sustained and stable PC control.
⑤ If you have an environment that you can connect to the LAN, you can check the image from the outside
⑥ Attract images and related information taken with i-BIS and create a movie image (AVI MOV MP 4)
⑦ You can easily see the front and back images of the top and the bottom
⑧ Check the detailed image of the well and manually graph it in time series Display Score for the condition
i-BISspecification
Chamber
- External dimensions (not including protrusions): W527XD845XH380 (mm)
- Dish inside dimension: 9 of LifeGlobalGPS dishes (60mm)
- Humidification: no humidification
- Power supply: AC100-120V 5A, 200-240V 2.5A, 50-60Hz
- Weight: Approx 60 Kg
Temperature
- Temperature Control:PID Control
- Use ambient temperature: 24-28oC
- Control temperature range: 36-39oC
- Fluctuation range: ±0.3oC (at room temperature 26oC)
- Temp Uniformity: ±0.3℃(at room temperature 26oC)
- Main course: Silicon rubber heaters
XY stage
- Movement range: ±lOOmm (x axis) ±150 mm (V axis)
- Repeat positioning accuracy : ±10 um
- Maximum speed: 30 mm / sec
Z stage
- Movement range: ±5mm
- Repeat positioning accuracy : ±0.3 um
- Maximum speed:±5 mm / sec
Gas specification
Supply gas: CO2 gas, N 2 gas (AGM system ※ 1)
CO2:
- Sensor: infrared sensor
- Control system: Duty control
- Control range: 0 - 10.0%
- Fluctuation range:±0.3%
O2
- Sensor: ceramic oxygen sensor
- Control system: Duty control
- Control range: 2 - 18.0%
- Fluctuation range:±0.5%
Optical system
- Camera: CCD 1.3 million pixels
- Objective lens: 10X
- Light source: Red LED lighting
image
- Image storage format: JPEG
- Image storage method: NAS (Mirroring function with function 4 TB x 2)
External output
- Analogue output: temperature (chamber 1 x 9) CO2 concentration, O2 concentration SMP connector 18 pins (manufactured by JST)
- CLOSE at contact output alarm occurrence, CLOSE at power OFF, SMP connector 4 pin (manufactured by JST)
※ Each data of this product is measured by our standards.
|